Imaging retinal densitometry with a confocal scanning laser ophthalmoscope

article
We describe a novel use of the Scanning Laser Ophthalmoscope (SLO), viz. as an imaging retinal densitometer. In our SLO a helium-neon or an argon laser beam is moved in a raster pattern over the retina; the reflected light is descanned (confocal SLO) and collected by a photomultiplier. Images of the fundus subtending 22 by 18 deg are displayed on a TV monitor. Single frames taken with 514 nm light were stored in a computer in arrays of 256 by 256 pixels and density differences between dark adapted and bleached images were calculated. With a full bleach density differences of about 0.35 were found in the center of the fovea; at retinal eccentricities of 15–20 deg we found 0.15. After selective bleaching with 633 nm light substantial density differences were only seen in the foveal area. We conclude that the confocal SLO is a very suitable instrument for imaging fundus reflectometry.
TNO Identifier
6792
Source
Vision Research, 29(12), pp. 1825-1830.
Pages
1825-1830