Real-time characterization of chemical threat agent aerosols for improvement of inhalation studies
article
Objectives: Deliberate or accidental release of chemical treat agents in the aerosol form can cause an inhalation hazard. Since the relationship between aerosol properties and health hazards is poorly understood, research into the toxicological consequences of exposure to aerosols is needed. The aim of the present study was to improve the characterization of particles for inhalation studies.
Methods: Several aerosol measurement technologies were compared for their potential to physically and chemically characterize particles in the inhalation size range in real-time. For that purpose, we compared the performance of an aerodynamic particle sizer (APS), a scanning mobility particle sizer (SMPS) and an electrical low-pressure impactor (ELPI) in an experimental set-up in which particles were generated by a Collison nebulizer and subsequently delivered into a nose-only inhalation exposure system.
Results: We found that more than 95% of the number of particles, equating to more than 83% of the mass generated by the 6-jet Collison nebulizer, were below 0.5 mm. To characterize the entire size range, the APS as single detector has only limited value, therefore the addition of supplementary instrumentation such as the SMPS or the ELPI is required. After real-time measurements in the size range of 30nm to 10 mm, ex-situ chromatographic chemical analysis is essential for quantification of the delivered mass concentration.
Conclusions: In summary, the present work demonstrates the utility of the ELPI technology, in combination with off-line analysis, for characterizing aerosols with various size, shape, charge, and composition. This makes the aerosol generation and analysis suite described a promising tool for quantitative inhalation exposure studies.
Methods: Several aerosol measurement technologies were compared for their potential to physically and chemically characterize particles in the inhalation size range in real-time. For that purpose, we compared the performance of an aerodynamic particle sizer (APS), a scanning mobility particle sizer (SMPS) and an electrical low-pressure impactor (ELPI) in an experimental set-up in which particles were generated by a Collison nebulizer and subsequently delivered into a nose-only inhalation exposure system.
Results: We found that more than 95% of the number of particles, equating to more than 83% of the mass generated by the 6-jet Collison nebulizer, were below 0.5 mm. To characterize the entire size range, the APS as single detector has only limited value, therefore the addition of supplementary instrumentation such as the SMPS or the ELPI is required. After real-time measurements in the size range of 30nm to 10 mm, ex-situ chromatographic chemical analysis is essential for quantification of the delivered mass concentration.
Conclusions: In summary, the present work demonstrates the utility of the ELPI technology, in combination with off-line analysis, for characterizing aerosols with various size, shape, charge, and composition. This makes the aerosol generation and analysis suite described a promising tool for quantitative inhalation exposure studies.