Galvanic detection of sulfur dioxide in ambient air at trace levels by anodic oxidation
article
A continuous method for the measurement of SO2 in ambient air at trace levels is described. The principle of detection is based on the anodic oxidation of SO2 in a galvanic cell. A differential measuring technique with a cell with two anodes and one cathode is used; background and noise current are low and stable. A feature of the measuring system is that the air sample is passed through a porous anode against a head of electrolyte on the other side of the anode. The minimum values for the time constant and the lower detection limit were 3 seconds and about 3 μg/m3, respectively. The selectivity of the method was compared with those of other techniques in field testing. Trace concentrations measured by the galvanic method in the range from 12 to 135 μg/m3 were in good agreement with those of the flame photometric and the West-Gaeke method. With a Pt/Ni2B/graphite anode the mean oxidation efficiency was 94 ± 2% on the basis of 2 faraday/mol of SO2 within the tested concentration range of 50 to 560 μg/m3. During a field test no decrease in the anode activity was observed over a period of ten months. Examples are given for the galvanic detection of other compounds using porous measuring electrodes. Chemicals/CAS: sulfur dioxide, 7446-09-5
TNO Identifier
228366
ISSN
08940630
Source
Journal of the Air Pollution Control Association, 28(2), pp. 138-142.
Pages
138-142