Print Email Facebook Twitter An improved method for the detection of α2-antiplasmin activity on the automated coagulation analyzer MLA Electra 1000C Title An improved method for the detection of α2-antiplasmin activity on the automated coagulation analyzer MLA Electra 1000C Author Meijer, P. Westerveld, W. Huisman, L.G.M. Gaubius Laboratory TNO Preventie en Gezondheid Publication year 1994 Abstract Comparison of the results for α2-antiplasmin activity of plasma samples (n = 61, range 0-125%) obtained by the 'immediate plasmin inhibition test' (IPIT) and by the automated method for the MLA Electra 1000C of Dade® shows a good correlation (R = 0.935) but a systematic disagreement especially relevant in the range below 40%. An improved method for the detection of α2-antiplasmin activity on the automated coagulation analyzer MLA Electra 1000C was developed. A linear calibration curve up to 150% was observed. A dilution curve of normal pooled plasma with plasma of an α2-antiplasmin deficient patient shows a linear curve (y = x + 7). Subject Biologyα2-antiplasminautomated methodplasminalpha 2 antiplasminplasminassayautoanalyzerautomationconference papercontrolled studyfibrinolysishumanmajor clinical studyplasma To reference this document use: http://resolver.tudelft.nl/uuid:78c7fe5e-28a8-4453-afcf-410153bcc753 DOI https://doi.org/10.1016/0268-9499(94)90284-4 TNO identifier 232684 ISSN 0268-9499 Source Fibrinolysis, 8 (SUPPL. 2), 160-162 Document type article Files To receive the publication files, please send an e-mail request to TNO Library.