Print Email Facebook Twitter The use of Stabilyte® plasma may cause changes in pH in the assay of some fibrinolysis analytes and might effect results Title The use of Stabilyte® plasma may cause changes in pH in the assay of some fibrinolysis analytes and might effect results Author Meijer, P. van der Ham, F. Kluft, C. Gaubius Laboratory TNO Preventie en Gezondheid Publication year 1996 Abstract The use of Stabilyte® plasma in the Coaliza® t-PA, Coaliza® PAI-1 and Enzygnost® PAP micro was evaluated. For PAI-1 antigen no, for t-PA antigen only small differences were observed between measurements in citrate and Stabilyte® plasma. For PAP-complexes significant lower (72%) results were obtained when Stabilyte® plasma was applied. This is the result of a lower pH during the first step of the test in comparison with citrated plasma. Subject Biologyassayclinical articleconference papercontrolled studyfibrinolysishumannormal humanphplasmapriority journalstandardization To reference this document use: http://resolver.tudelft.nl/uuid:35c5673f-9b77-4103-897c-870fcb6852b6 DOI https://doi.org/10.1016/s0268-9499(96)80075-0 TNO identifier 233526 ISSN 0268-9499 Source Fibrinolysis, 10 (SUPPL. 2), 155-157 Document type article Files To receive the publication files, please send an e-mail request to TNO Library.